• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 10X140MM; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 10X140MM; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: 51-104100 tprlc 133 t1 pps ho 10x140mm lot number 7204979.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2023 - 01546.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.H3 other text : device location is unknown.
 
Event or Problem Description
It was reported that during surgery, the doctor placed the stem and the implant sat proud.The surgeon tried another implant of the same size and it also sat proud.The procedure was completed using a smaller size stem.There is no additional information available at the time of this report.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4.D4: expiration date and udi.D9.G3.G6.H2.H6: component code: mechanical (g04) - stem.H10.Visual examination of the returned product identified there was no visible damage to either of the devices.The height of the device was checked per the print in with both devices are conforming.The height gage was used to determine the spots to measure the stem thickness at three areas along with biomet spec.All of the measurements were found within tolerance of the print.Complaint confirmed based on returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.No device problem was found.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS HO 10X140MM
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17317556
Report Number0001825034-2023-01545
Device Sequence Number3995978
Product Code LZO
UDI-Device Identifier00880304491830
UDI-Public(01)00880304491830(17)310524(10)7026901
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-104100
Device Lot Number7026901
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Supplement Date Received by Manufacturer07/26/2023
Initial Report FDA Received Date07/13/2023
Supplement Report FDA Received Date08/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
-
-