Catalog Number 60-6085-202A
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of the customer, that the 60-6085-202a, vcare 200a, large was being used during a hysterectomy procedure on (b)(6) 2023 when it was reported ¿vcare handle came loose and was not able to manipulate the uterus properly.¿ the procedure was completed without the use of an alternate device.There was no report of injury, medical intervention, or extended hospitalization to the patient or user.This report is being raised on the reported malfunction with potential for injury upon reoccurrence.
 
Event or Problem Description
The sales representative reported on behalf of the customer, that the 60-6085-202a, vcare 200a ¿ large was being used during a hysterectomy procedure on (b)(6) 2023 when it was reported "vcare handle came loose and was not able to manipulate the uterus properly".The procedure was completed without the use of an alternate device.There was no report of injury, medical intervention, or extended hospitalization to the patient or user.This report is being raised on the reported malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device will not be returned for evaluation and no photographic evidence was provided therefore the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised that prior to device removal, ensure the locking mechanism is released via the thumbscrew and swipe a finger around the edge of the vaginal cup to separate the tissue from the cup to prevent tissue damage.Do not use excessive force upon device removal to avoid traumatizing the vaginal canal and/or component detachment.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCARE 200A - LARGE
Common Device Name
CANNULA, MANIPULATOR/INJECTOR, UTERINE
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key17319649
Report Number1320894-2023-00138
Device Sequence Number3996374
Product Code LKF
UDI-Device Identifier10653405003809
UDI-Public(01)10653405003809(17)240814(10)202208151
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142716
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/14/2024
Device Catalogue Number60-6085-202A
Device Lot Number202208151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2023
Supplement Date Received by Manufacturer07/14/2023
Initial Report FDA Received Date07/13/2023
Supplement Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient EthnicityNon Hispanic
-
-