| Model Number |
IPN915197 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that the screw of the handle was found loose when checking it at the hospital supply room.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The sample was returned and sent to the manufacturing site (tecomet) for investigation.Tecomet reports "the dhr for the returned instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, inc.Kenosha wi facility as part of a 50-pc.Lot in september of 2021.It was found that this order was made from the correct materials and components and all approved processes were followed.Evaluation of the returned instrument as received shows that the endo handle pivot screw (n00186) is easily backed out of its threading which signifies that the staking operation that holds the screw in place in between the g00436m handle has failed.Further evaluation shows that the staking operation has been performed at the time of manufacture.We are unable to determine what cause the screw to become loose from its staking operation since we do not know how this instrument has been handled out in the field since its date of production.Mishandling of this device at the end user's facility is suspected.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.".
|
| |
|
Event or Problem Description
|
|
It was reported that the screw of the handle was found loose when checking it at the hospital supply room.No patient involvement.
|
| |
|
Search Alerts/Recalls
|