ETHICON ENDO-SURGERY, LLC. GOLD RELOAD FOR ECHELON 45; STAPLE, IMPLANTABLE Back to Search Results
Model Number ECR45D
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 7/13/2023 d4: batch # unk a manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 8/10/2023.D4: batch # 848a21.Investigation summary : the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload.  visual analysis of the returned sample determined that one ecr45d reload was received partially fired 1/10.The returned reload was reset and loaded into a test device.The device was tested for functionality in the straight position and achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.  it is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.Please reference the instruction for use for more information.  device history review: a manufacturing record evaluation was performed for the finished device batch number 848a21, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOLD RELOAD FOR ECHELON 45
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key17321416
Report Number3005075853-2023-04931
Device Sequence Number20561037
Product Code GDW
UDI-Device Identifier10705036001737
UDI-Public10705036001737
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberECR45D
Device Catalogue NumberECR45D
Device Lot Number954A81
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/20/2023
Supplement Date Received by Manufacturer07/18/2023
Initial Report FDA Received Date07/13/2023
Supplement Report FDA Received Date08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-