| Model Number |
6000 SERIES |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
Abrasion (1689)
|
| Date of Event |
06/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The user facility engineering team inspected the endo drying cabinet and found several unsmoothed edges.The unit was removed from service and instruction was given to the user facility personnel to avoid using the cabinet until repairs have been made.A steris service technician arrived onsite following the reported event to inspect the drying and storage cabinet.The technician found several sharp edges within the unit potentially causing the reported event to occur.The shelving within the drying and storage cabinet will be replaced, and all sharp edges will be smoothed out.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that an employee obtained a cut injury while retrieving equipment from their reliance 6000 series endoscope storage cabinet.Medical treatment was sought and administered (stitches).
|
| |
|
Additional Manufacturer Narrative
|
|
The technician replaced the shelving and all sharp edges on the unit were smoothed out.The unit was tested, confirmed to be operating according to specifications, and returned to service.No additional issues have been reported.
|
| |
|