The reason for this revision surgery, the agent reported "(anatomic glenoid poly loosening.Surgeon requested me to open parts in new dt to perform anatomic to reverse conversion as the stem was viable though he changed his mind and did not implant already opened implants.Took the turon stem out and did not implant turon-rsp conversion shell and std shell poly.Zimmer reverse shoulder was then used to complete the revision as this was his backup plan)".The actual length of in-vivo for the item listed is unknown as the original surgery date was not provided or could not be established.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loosening.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical poor bone density, inadequate soft tissue support, patient activities or trauma.
|