| Model Number |
90495 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hold for jf 7.14.It was reported that during an endovascular procedure the subject balloon was not deflating before removing it from internal carotid artery ica.Hence doctor had to pull out with out balloon deflation.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 other text : the device is not available to the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: manufacturing date ¿ added.D4: expiration date - added.There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual testing as well as functional testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information received indicated that the device was prepared for use as per the directions for use.There was no damage noted to the packaging prior to opening the packaging and the device was confirmed to be in good condition during preparation/prior to use on the patient.Continuous flush was set up and maintained throughout the clinical procedure.The balloon was successfully deflated during preparation but it was reported that 'the balloon was not deflating before removing it from ica.Hence doctor has to pull out with out balloon deflation.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint.
|
| |
|
Event or Problem Description
|
|
It was reported that during an endovascular procedure the subject balloon was not deflating before removing it from internal carotid artery ica.Hence doctor had to pull out with out balloon deflation.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|
| |
|