|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the control body fluid damage, the lcb (light carrying bundle) fluid damage, the light guide cable fluid damage, the us connector cable (long) fluid damage, the us connector/junction cable (short) fluid damage, the ultrasound connector body fluid damage, the lg connector fluid damage, the objective lens cloudy (not clear view), and the operation channel (primary) leak; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|