• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,3.7,16,MTX,MG; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,3.7,16,MTX,MG; DENTAL IMPLANT Back to Search Results
Catalog Number TSVTB16
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age at time of the event unknown / not provided.A3: gender unknown / not provided.A4: patient weight unknown / not provided.
 
Event or Problem Description
Doctor reported that the collar implant fractured and was removed.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Zimvie received one (1) tsvtb16, (imp,tsv,3.7,16,mtx,mg) for evaluation.Visual evaluation was performed, a fracture has been identified on the implants collar.Dhr review was completed for the subject lot number 1220716.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1220716 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : implant.Based on the investigation and risk management file review, the most likely root cause determined from the investigation were missing or confusing instructions for use and clinician selects implant that is unable to resist long-term occlusal loading.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the collar.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,3.7,16,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17323078
Report Number0002023141-2023-01909
Device Sequence Number12134329
Product Code DZE
UDI-Device Identifier00889024019959
UDI-Public(01)00889024019959(17)231009(10)1220716
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K101977
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/08/2023
Device Catalogue NumberTSVTB16
Device Lot Number1220716
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/14/2023
Supplement Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/08/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-