• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number MOSAIQ
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported an instance of incorrect dose recording.
 
Additional Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Additional Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported an instance of incorrect dose recording.From the mosaiq logs the investigation found that the sequencer crashed mid treatment, so the treatment was not recorded.Mosaiq received the sro file, dicom rt image for port during integrate option as well as the dicom beam records for two fields which were treated in full.However, these files were not processed because of the vmi's error, "vmicheckerrormsg: failed to invoke c-store-rq.Reason: opening association failed with impac_dcm_scp".The error indicated that there were some problems communicating with the dcmwin app.Based on available information there was no mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Common Device Name
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA INC
100 mathilda place
5th floor
sunnyvale CA 94086
Manufacturer (Section G)
ELEKTA INC
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key17323228
Report Number2950347-2023-00006
Device Sequence Number5641865
Product Code IYE
UDI-Device Identifier00858164002091
UDI-Public(01)00858164002091(10)2.64.235
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K141572
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2023
Supplement Date Received by Manufacturer06/16/2023
Initial Report FDA Received Date07/14/2023
Supplement Report FDA Received Date11/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-