Model Number ELI380-ACX21
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.During troubleshooting the hrc technician and the customer's biomed determined that the event had been caused by the mainboard and radio board not working as intended and needing to be replaced.Based on this information, no further action is required at this time.If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ecg¿s.This failed connection may result in a delay of patient treatment which may cause or contribute to a serious injury or death.Therefore, hillrom is reporting this alleged connection failure as a product malfunction.
 
Event or Problem Description
The customer reported that the eli380 intermittently loses wlan connectivity.The network card is not recognized.After a while it turns on but then says it can¿t read the worklist.This incident was captured under hillrom complaint ref (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELI380 LCD AM12 WLAN USB DICOM AHA BAN
Common Device Name
ELECTROCARDIOGRAPH
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
keighley crosthwaite
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key17324953
Report Number2183461-2023-00015
Device Sequence Number12349195
Product Code DPS
UDI-Device Identifier00812345024562
UDI-Public812345024562
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Model NumberELI380-ACX21
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2023
Initial Report FDA Received Date07/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-