| Model Number |
461503 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reports: the purple end tip popped off from the inserted 5french 36 feeding tube being used on a baby for feeds.The tube was removed immediately with no known concerns to patient.Lot numbers that were available at the unit are lot #s 2231310564 or #2215831164.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4 - lot number: the lot number was updated per additional information provided by the customer on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record file was reviewed, and no discrepancy was found according to the reported condition.The analyzed sample showed a detachment of the purple connector; in order to determine a root cause, a gemba walk was done at the manufacturing facility focusing on the manufacturing equipment and process; it was determined as a potential root cause was that the drying time was below specification.As corrective/preventative action, personnel production received an awareness notification of the reported complaint.
|
| |
|