| Model Number |
26160SHW |
| Medical Device Problem Codes |
Crack (1135); Material Fragmentation (1261); Pitted (1460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id:(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that fragment was identified in the uterus when viewed on the monitor during the cervical procedure with the 26160shw(zp02).The procedure was delayed to identify and remove the debris; however, the debris was very fine and could not be removed.There has been no impact on the patient so far.
|
| |
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The claimed item 26160 shw with the date of manufacture december 2020 was found to be damaged at the distal end.The article has cracks in the material and a notch in the fork.A check of the inspection lot shows that no defects were found during the inspection.The initial stage of pitting and the notch developing into a crack.The piece missing from the notch is less than a millimetre in size, which may indicate that this piece has rusted away due to corrosion (as pitting can be seen at the edges), or it may have come loose during surgery.Therefore, as described in the instructions for use the item should be checked for function and damage before each use.Otherwise, pieces may come loose due to damage or, as in this case, corrosion.Internal karl storz reference number: (b)(4).
|
| |
|