• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG SCISSORS, 5 FR.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG SCISSORS, 5 FR. Back to Search Results
Model Number 26160SHW
Medical Device Problem Codes Crack (1135); Material Fragmentation (1261); Pitted (1460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id:(b)(4).
 
Event or Problem Description
It was reported that fragment was identified in the uterus when viewed on the monitor during the cervical procedure with the 26160shw(zp02).The procedure was delayed to identify and remove the debris; however, the debris was very fine and could not be removed.There has been no impact on the patient so far.
 
Additional Manufacturer Narrative
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
The claimed item 26160 shw with the date of manufacture december 2020 was found to be damaged at the distal end.The article has cracks in the material and a notch in the fork.A check of the inspection lot shows that no defects were found during the inspection.The initial stage of pitting and the notch developing into a crack.The piece missing from the notch is less than a millimetre in size, which may indicate that this piece has rusted away due to corrosion (as pitting can be seen at the edges), or it may have come loose during surgery.Therefore, as described in the instructions for use the item should be checked for function and damage before each use.Otherwise, pieces may come loose due to damage or, as in this case, corrosion.Internal karl storz reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCISSORS, 5 FR.
Common Device Name
SCISSORS, 5 FR.
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave.
el segundo, CA 90245
MDR Report Key17326738
Report Number9610617-2023-00159
Device Sequence Number17733913
Product Code GCJ
UDI-Device Identifier04048551094655
UDI-Public4048551094655
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K935071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26160SHW
Device Catalogue Number26160SHW
Device Lot NumberZP02
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2023
Supplement Date Received by Manufacturer07/25/2023
10/11/2023
Initial Report FDA Received Date07/14/2023
Supplement Report FDA Received Date08/10/2023
10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-