| Catalog Number |
TSVM6H10 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Hypersensitivity/Allergic reaction (1907)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint (b)(4).G4: additional 510(k) number is k101880.
|
| |
|
Event or Problem Description
|
|
Patient returned to clinic after being evaluated by may clinic and recent specialized titanium testing which confines, he's allergic to the titanium.Tooth #'s 2, 23 and 26.Symptoms as a result of the event: pain, inflammation, sensitivity and metallic taste.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number.H2: if follow-up, what type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: type of investigation.H6: investigation findings.H6: investigation conclusions.H10: additional narratives/data.Zimvie received one implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, no apparent malfunction was identified with the implant that would contribute to the reported event.Markings due to the removal process.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.Therefore, based on the available information, a device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|