BOSTON SCIENTIFIC CORPORATION SYNERGY MEGATRON; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493942832400
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that removal difficulties and catheter entrapment occurred.The 70% stenosed target lesion was located in the non-tortuous and mildly calcified left main coronary artery.A 4.00 x 32mm synergy megatron drug-eluting stent was advanced for treatment.The stent balloon was inflated at 14 atmospheres for roughly 15 seconds and the stent was deployed.After waiting 20-30 seconds for the balloon to fully deflate, removal difficulties were encountered.The stent balloon became stuck in the proximal portion of the guide catheter and the guidewire appeared to be stuck to the balloon.The balloon was still stuck even after pulling back pressure/vacuum on the indeflator a second time.The stent balloon, wire, and guide catheter had to be removed as one unit.The procedure was completed with no patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY MEGATRON
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17329879
Report Number2124215-2023-34701
Device Sequence Number5487521
Product Code NIQ
UDI-Device Identifier08714729985785
UDI-Public08714729985785
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2024
Device Model NumberH7493942832400
Device Lot Number0030832797
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2023
Initial Report FDA Received Date07/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER: MDT EBU 4.0 7F
Patient SexUnknown
-
-