| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user was explanted on (b)(6) 2023.As per explant report form information, the electrode array was found only partially inserted.
|
| |
|
Event or Problem Description
|
|
The user was explanted on (b)(6) 2023.As per explant report form information, the electrode array was found only partially inserted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received from the field the device was explanted due to a post-operative migration of the active electrode out of cochlea as confirmed by diagnostic imaging.This is a final report.
|
| |
|