| Catalog Number |
8065752082 |
| Medical Device Problem Codes |
Complete Blockage (1094); Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that an abnormal sound was heard, and aspiration failure occurred during the surgery.The surgery was completed without product replacement.There was no patient harm.The procedure type was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A used intrepid fluidics management system (fms) was visually inspected, and no obvious defects were detected.The sample was tested using a console.The fms primed and tuned with the ultrasonic handpiece, the 0.9-millimeter (mm) aspiration bypass system (abs) tip and infusion sleeve from the lab stock successfully.No abnormal sounds happened from the pump elastomer.Cassette max vacuum and aspiration flow rate, and irrigation flow rate were measured and met specifications.No problem was found with the returned cassette fluidics.The root cause of the customer's complaint could not be established; the returned sample functioned per specification.This complaint has been reviewed and it is determined that no further actions will be pursed at this time as the sample met specifications.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|