| Model Number |
192-N251S |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each additional lot number is as follows: d.4.Medical device lot #: 2188470, d.4.Medical device expiration date: 30-jun-2027, h.4.Device manufacture date: 07-jul-2022, d.4.Medical device lot #: 2153548, d.4.Medical device expiration date: 31-may-2027, h.4.Device manufacture date: 02-jun-2022, d.4.Medical device lot #: 2188472, d.4.Medical device expiration date: 30-jun-2027, h.4.Device manufacture date: 07-jul-20222.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 of the bd safetyglide¿ needle experienced leakage.The following information was provided by the initial reporter: leaks at the hub.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no samples or photos received by our quality team for evaluation.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.During review of the process, process variations during the needle lubricant application can create clogged needles.Corrective and preventative actions have been implemented.Several quality initiatives have been implemented on our manufacturing line to ensure that the needle lubricant application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the needle lubricant is applied uniformly to the needle.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 of the bd safetyglide¿ needle experienced leakage.The following information was provided by the initial reporter: leaks at the hub.
|
| |
|
Search Alerts/Recalls
|