W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT351414
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 06/22/2023
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2016, the patient was treated for an abdominal aortic aneurysm.The physician implanted a gore® excluder® aaa endoprosthesis main body and on gore® excluder® aaa endoprosthesis.The patient tolerated the procedure.On an unknown date, the patient received a ct scan related to a kidney function issue.At that time, it was seen that the main body device had migrated into the aneurysm sac, causing a type ia endoleak.The physician extended proximally with a gore® excluder® aaa endoprosthesis, which sealed the endoleak.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H.6.Type of investigation code b21 updated to code b14 with completion of phr review.H.6.Investigation findings: code c21 updated to code c19 with completion of phr review.According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to, endoleak and component migration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
greg rawlings
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key17333673
Report Number3007284313-2023-02639
Device Sequence Number6647262
Product Code MIH
UDI-Device Identifier00733132618477
UDI-Public00733132618477
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/19/2020
Device Model NumberRLT351414
Device Catalogue NumberRLT351414
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2023
Supplement Date Received by Manufacturer06/22/2023
Initial Report FDA Received Date07/17/2023
Supplement Report FDA Received Date07/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexMale
Patient Weight69 KG
-
-