• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS SMART PORT; PORT & CATHETER,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS SMART PORT; PORT & CATHETER, Back to Search Results
Model Number CT80STPD
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Foreign Body In Patient (2687)
Date of Event 07/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
Hensley legal group, pc., a law group, notified angiodynamics of an incident regarding a smartport from a single titanium ct vtx port w poly cath kit.The device was implanted by (b)(6) at (b)(6)., for the purpose of ongoing medical treatment on (b)(6) 2021.On or about (b)(6)2021, the smartport [catheter] fractured, causing severe injury to plaintiff.On or about (b)(6) 2021, plaintiff underwent surgery with (b)(6) at iu (b)(6), to remove the smartport and the fractured catheter.As a result of having the smartport implanted, plaintiff has reportedly experienced significant mental and physical pain and suffering, has sustained permanent injury, permanent and substantial physical deformity, has undergone and will undergo corrective surgery or surgeries, and has suffered financial or economic loss, including, but not limited to, obligations for medical services and expenses, and present and future lost wages.
 
Additional Manufacturer Narrative
It was reported that the disposable device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of the smartport [catheter] fractured, causing severe injury to plaintiff.Cannot be confirmed; no product samples were returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A potential root cause for the reported catheter fractured (and leaked) failure mode is pinch-off syndrome; which is cautioned in the device dfu as potential complication.Labeling review: the instructions for use, which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome: pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Note: if infusion or aspiration is successful upon lifting arm above the head and turning the head, consider pinch-off syndrome as a possible cause.The line should be radiologically evaluated if pinch-off syndrome is suspected.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT
Common Device Name
PORT & CATHETER,
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key17334291
Report Number1317056-2023-00090
Device Sequence Number7614130
Product Code LJT
UDI-Device IdentifierH787CT80STPD0
UDI-PublicH787CT80STPD0
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K101017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberCT80STPD
Device Catalogue NumberCT80STPD
Device Lot Number5645791
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer07/20/2023
Initial Report FDA Received Date07/17/2023
Supplement Report FDA Received Date07/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-