• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 6900PTFX31
Medical Device Problem Codes Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 05/24/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was learned from implant patient registry that a patient with a 31mm 6900ptfx pericardial mitral valve underwent a redo mitral valve replacement procedure after an implant duration of 7 years, 1 month due to unknown reasons.The explanted valve was replaced with a 31mm 11400m valve.The patient was noted to be in recovery post procedure.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: updated: b4, b5, g3, g6, h2, h6.
 
Event or Problem Description
It was learned from implant patient registry that a patient with a 31mm 6900ptfx pericardial mitral valve underwent a redo mitral valve replacement procedure after an implant duration of 7 years, 1 month due to mitral stenosis.The patient initially presented with heart failure and shortness of breath.The explanted valve was replaced with a 31mm 11400m valve.The patient was noted to be in stable condition in the icu post procedure.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: the device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key17334842
Report Number2015691-2023-14637
Device Sequence Number15304938
Product Code DYE
UDI-Device Identifier00690103163612
UDI-Public(01)00690103163612(17)190221
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date02/21/2019
Device Model Number6900PTFX31
Device Catalogue Number6900PTFX31
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2023
Supplement Date Received by Manufacturer07/25/2023
08/24/2023
Initial Report FDA Received Date07/17/2023
Supplement Report FDA Received Date08/18/2023
09/06/2023
Date Device Manufactured02/22/2015
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age63 YR
Patient SexFemale
-
-