|
There was an allegation of questionable gen.2 ise indirect for na results for 10 patient samples on a cobas e 411 immunoassay analyzer.The customer was having calibration issues with the na electrode and replaced it before repeating the samples.For patient 1, the initial na result was 134 mmol/l.The repeat result was 139.28 mmol/l.For patient 2, the initial na result was 133 mmol/l.The repeat result was 141 mmol/l.For patient 3, the initial na result was 120 mmol/l.The repeat result was 135 mmol/l.For patient 4, the initial na result was 130 mmol/l.The repeat result was 136 mmol/l.For patient 5, the initial na result was 133 mmol/l.The repeat result was 140 mmol/l.For patient 6, the initial na result was 134 mmol/l.The repeat result was 141 mmol/l.For patient 7, the initial na result was 135 mmol/l.The repeat result was 143 mmol/l.For patient 8, the initial na result was 134 mmol/l.The repeat result was 141 mmol/l.For patient 9, the initial na result was 133 mmol/l.The repeat result was 141 mmol/l.For patient 10, the initial na result was 134 mmol/l.The repeat result was 140 mmol/l.No questionable results were reported outside of the laboratory.The repeat results were deemed correct.
|