• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRECISION SPINE, INC REFORM TI CT MODULAR MIS PEDICLE SCREW; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRECISION SPINE, INC REFORM TI CT MODULAR MIS PEDICLE SCREW; SCREWDRIVER Back to Search Results
Model Number 64-SP-1700
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3 other: evaluation in process but not yet complete.Upon completion of evaluation, a follow up report will be submitted.This report is number 1 of 4 mdrs filed for the same event (reference 3005739886-2023- 00033).
 
Event or Problem Description
It was reported that a procedure was performed on (b)(6) 2023, utilizing the reform ti mis ct pedicle screw system.Upon screw insertion the tip of two polyaxial driver assembly, cannulated reform mis system (64-sp-1700), and one adjustment driver, reform mis pedicle system (64-sp-1700) broke.A ø7.5mm x 45mm reform ti modular cannulated screw (64-sp-7545) also sheared at the neck.The sheared screw was removed and replaced.No broken tips were left in the patient, but there was a 45-55 minute delay to the procedure due to the malfunctions.
 
Additional Manufacturer Narrative
H3 device evaluation: screwdriver exhibits two damaged areas.It's drive shaft hexalobe is twisted and a fracture also occurred at the 1/4" square's pinned interface.No signs of failure due to combination loading were observed.The failure is consistent with failure due to high torque application since the failure plane is normal to the driver's longitudinal axis.Crack initiation from prior high torque application cannot be ruled out as a cause for the failure observed in the given procedure.A review of the device history records (manufacturing and inspection records), and a review of failures for this lot did not bring any manufacturing related anomalies to light.The cause for the failures is undetermined but is believed to be due to high torque application.No corrective actions are being recommended.This report is number 1 of 4 mdrs filed for the same event (reference 3005739886-2023-00030-1 / 00033-1).
 
Event or Problem Description
It was reported that a procedure was performed on (b)(6) 2023, utilizing the reform ti mis ct pedicle screw system.Upon screw insertion the tip of two polyaxial driver assembly, cannulated reform mis system (64-sp-1700) and one adjustment drive, reform mis pedicle system (64-sp-0601) broke.A 07.5mm x 45mm reform ti modular cannulated screw (64-sk-7545) also sheared at the neck.The sheared screw was removed and replaced.No broken tips were left in the patient, but there was a 45-55 minute delay to the procedure due to the malfunctions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFORM TI CT MODULAR MIS PEDICLE SCREW
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer (Section G)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer Contact
mike dawson
2050 executive drive
pearl, MS 39208
6014204244
MDR Report Key17337171
Report Number3005739886-2023-00030
Device Sequence Number6617407
Product Code HXX
UDI-Device Identifier00840019942454
UDI-Public00840019942454
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number64-SP-1700
Device Catalogue Number64-SP-1700
Device Lot Number00302PT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2023
Supplement Date Received by Manufacturer06/22/2023
Initial Report FDA Received Date07/17/2023
Supplement Report FDA Received Date10/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-