• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SMPX545
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: right capsular contracture; baker grade iv.D6b explantation date: (b)(6) 2023.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 45-year-old hispanic female patient underwent primary breast augmentation with a 545cc mentor memorygel xtra breast implant and experienced capsular contracture; baker grade iv on her right side postoperatively; diagnosed after physical exam.The patient has consulted with the healthcare professional.As a result, the device will be explanted on (b)(6) 2023.
 
Additional Manufacturer Narrative
On march 18, 2024, mentor became aware that the date of surgery is (b)(6) 2024.Hence, following fields have been updated on this form: is required intervention has been selected under section b; field b2 fields d6b for explantation date is (b)(6) 2024.Field h6 health effect - impact code ¿device explantation¿ has been added field h6 type of investigation has been updated to "device not returned" reason for device explant and/or reoperation: right capsular contracture; baker grade iv since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2024, , mentor became aware that the surgery has been postponed.New surgery date has not been scheduled.Hence, following fields have been updated on this form: - is required intervention has been de-selected under section b; field b2 - explantation date has been removed from fields d6b.- field h6 health effect - impact code has been updated to ¿recognized device or procedural complication¿ - field h6 type of investigation has been updated to "device not accessible for testing" reason for device explant and/or reoperation: as of now there is no information regarding the explant or reoperation.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On may 13, 2024, the mentor failure analysis lab received the device for evaluation.On may 17, 2024, device evaluation was completed as follows: mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the smooth mod + prof xtra 545cc returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.  as part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On may 30, 2024, mentor became aware that the patient underwent right side replacement surgery with a new 545cc mentor memorygel xtra breast implant on (b)(6) 2024.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key17339247
Report Number1645337-2023-08305
Device Sequence Number17868281
Product Code FTR
UDI-Device Identifier00081317025115
UDI-Public00081317025115
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSMPX545
Device Catalogue NumberSMPX545
Device Lot Number9703508
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/03/2023
Supplement Date Received by Manufacturer02/21/2024
03/18/2024
05/13/2024
05/30/2024
Initial Report FDA Received Date07/17/2023
Supplement Report FDA Received Date03/10/2024
03/20/2024
05/17/2024
05/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
Patient EthnicityHispanic
Patient RaceWhite
-
-