• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) SUSTAIN XL DC; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) SUSTAIN XL DC; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2134
Medical Device Problem Codes Loose or Intermittent Connection (1371); No Pacing (3268)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented for an implant procedure.It was noted that when the lead was connected to the the pacemaker, the screw was unable to be tightened and there was no output.The pacemaker was not used and replaced during procedure.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of pacing problem was not confirmed, and the reported event of setscrew issue was confirmed.The device was above elective replacement indicator (eri) level upon receipt.Both setscrews were able to tighten upon receipt.Analysis found that a setscrew insets was stripped due to material inside the inset preventing full wrench insertion of wrench, this is consistent with user error, which resulted in the reported event of the setscrew.Further analysis performed, including output verification and inspection of header and connectors showed normal device characteristics with no anomalies contributing to reported event of pacing problem.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.A device history record (dhr) review was performed, and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUSTAIN XL DC
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17339729
Report Number2017865-2023-36101
Device Sequence Number17868181
Product Code NVZ
UDI-Device Identifier05414734504720
UDI-Public05414734504720
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberPM2134
Device Catalogue NumberPM2134
Device Lot NumberP000137580
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2023
Supplement Date Received by Manufacturer09/25/2023
Initial Report FDA Received Date07/17/2023
Supplement Report FDA Received Date09/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LEAD
Patient SexUnknown
-
-