MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG RT MD SIZE 3 PMA; KNEE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG RT MD SIZE 3 PMA; KNEE PROSTHESIS Back to Search Results
Model Number 159575
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Associated products: oxf uni tib tray sz c rm pma; item# 154723; lot# 3645306 oxford uni twin-peg femoral md cocr med; item# 166942; lot# j3645657 foreign: japan.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that approximately seven years post initial implantation, the patient underwent a revision surgery for a knee bearing dislocation after a fall.Attempts have been performed and no further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG RT MD SIZE 3 PMA
Common Device Name
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17340653
Report Number3002806535-2023-00250
Device Sequence Number6912994
Product Code NRA
UDI-Device Identifier05019279786213
UDI-Public(01)05019279786213(17)200825(10)3621324
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/25/2020
Device Model Number159575
Device Catalogue Number159575
Device Lot Number3621324
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2023
Initial Report FDA Received Date07/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexPrefer Not To Disclose
-
-