• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. MEGAPOWER 1000 A GENERATOR; MEGA POWER 1000 A, GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. MEGAPOWER 1000 A GENERATOR; MEGA POWER 1000 A, GENERATOR Back to Search Results
Model Number 1000
Medical Device Problem Codes Arcing of Electrodes (2289); Human-Device Interface Problem (2949)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 06/30/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Serial number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what is the severity of the burn? (please see degrees of burns below and choose one) o first degree burns are minor burns on the first layer of skin.The skin looks dry, redness, and may be swelling; no penetration or blisters o second degree burn looks wet or moist.The burn site appears red, blistered, and may be swollen and painful o third degree burn the burn site looks deep, whitening or blackened and charred what medical intervention was used to treat the burn? (such as salve or stitches) besides the burn, did the patient experience any adverse consequence due to the issue? are there any anticipated long-term effects from the burn or injury? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure and before the incision; combustion occurs with subsequent fire, where they refer to disposable surgical fields being burned, upon evaluation of the patient, lesions are evidenced in the lateral region of the hip and right thigh, loss of hair is evidenced on the lower limbs, scrotum, and crotch secondary to the burn , redness is observed in the abdominal region, the position of the plate in the right shoulder blade is verified, which is connected to the console, the plastic surgeon is notified for evaluation and the surgical procedure is deferred it is important to highlight that the information received by the ward manager is that the megadyne generator was used in combination with medtronic return electrode technology.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 8/14/2023.Investigation summary: per service manual operational and diagnostic analysis did not confirm the reported unintended thermal injury or the flash flame fire.The unit passed all functional tests and is fully operational.No further investigation will be conducted on this complaint.The device history records were reviewed and certed by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGAPOWER 1000 A GENERATOR
Common Device Name
MEGA POWER 1000 A, GENERATOR
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key17341588
Report Number1721194-2023-00095
Device Sequence Number6894040
Product Code GEI
UDI-Device Identifier10614559103715
UDI-Public10614559103715
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K050579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1000
Device Catalogue Number1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer08/02/2023
Initial Report FDA Received Date07/18/2023
Supplement Report FDA Received Date08/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-