| Model Number |
N/A |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that outside the surgical environment the cart had a failed power inlet module.The heat from the fuses showed signs of charring and lite melting within the fuse compartment.There was no patient involvement.Due diligence is in process.No additional information has been received to date.No adverse events were reported as a result of this malfunction.
|
| |
|
Event or Problem Description
|
|
No additional event information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the unit was confirmed to have a failed power inlet module.The power inlet module fuse compartment was found to be charred and showed signs of melting.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Review of our reporting decision on fuses exhibiting signs of overheating.A review of all complaints for this product indicates that there has never been an event that has resulted in a death, serious injury, and/or actual fire when the fuse has blown or fuse drawer deformation.Therefore, zimmer biomet will not be reporting this failure mode going forward.
|
| |
|
Search Alerts/Recalls
|