| Model Number |
9066 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The probe itself was evaluated and bk medical did not find either a manufacturing defect or a malfunction.The investigation concluded that there is a high likelihood that the damaged flexible sheath is a result of mishandling of the transducer either 1) when sterilizing the transducer or 2) when retracting the transducer in a bend position or due to a combination of the two potential causes.All identified risk scenarios have been addressed in the product risk assessment and been mitigated to an acceptable level.The two potential causes are already addressed as cautions and warnings in the device user guide and user guide for care and cleaning.There is a low risk of any biological reactions to the patient.
|
| |
|
Event or Problem Description
|
|
Mdr report mw5113739.Received by bk medical 16-dec-2022.Event description: when inserting the bk medical laparoscopic probe the sheath covering the flexible tip tore and a piece fell in patient abdomen.
|
| |
|