• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number 363902
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: event 2: detailed inquiry description: the clinician was struggling with air in line during vancomycin infusion.No injury reported.
 
Additional Manufacturer Narrative
Event 2: this report has been identified as b.Braun medical internal report number (b)(4).One used sample with open packaging and four photographs were provided for evaluation.The sample and photographs were visually evaluated, and no defects were observed on the physical sample.In the photographs provided, microbubbles were visible in the tubing and the iv bag.In an attempt to replicate the reported defect of the air-in-line alarm on the pump, the physical sample was attached to the pump and then tested by running therapy while staggering the rate of flow throughout the therapy.No alarms occurred during the testing of the returned sample.The sample was also leak-tested with passing results.Based on the photos provided, the reported defect was confirmed.Although the photographs were confirmed and the air in line could not be replicated, a possible root cause of air in line alarms could be related to an incomplete priming of the iv line, infusion of cold solutions which may exhibit air bubbles coming out of solution as the fluid warms, or incomplete filling of the drip chamber during priming.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.In addition, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT®
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key17343371
Report Number2523676-2023-00354
Device Sequence Number6524072
Product Code FPA
UDI-Device Identifier04046964294730
UDI-Public(01)04046964294730
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number363902
Device Catalogue Number363902
Device Lot Number0061866016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2023
Supplement Date Received by Manufacturer06/28/2023
Initial Report FDA Received Date07/18/2023
Supplement Report FDA Received Date08/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-