• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23AGFN-756
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2023, a 23mm regent mechanical heart valve was chosen for implant for a patient undergoing aortic valve replacement (avr) procedure.It was reported after the valve was implanted, it was noticed the leaflets had "imperfect movement" and had difficulty moving freely.It was reported only the leaflet tester was used to test leaflet mobility and no other instrument or tool came in contact with the leaflets.A decision was made to explant and replace the 23mm valve with a 21mm regent mechanical heart valve.The replacement 21mm valve was successfully implanted in the patient.The 23mm valve was not tested for leaflet mobility after explant.It was reported there was a significant delay in the procedure due to this event.The patient remained hemodynamically stable throughout the procedure.The patient status was reported as stable.
 
Additional Manufacturer Narrative
An event of the valve leaflets being unable to open and close properly was reported.Morphological and hydrodynamic examination indicated the valve met abbott specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Hydrodynamic testing upon return to abbott and at the time of manufacturing indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17346653
Report Number2135147-2023-03085
Device Sequence Number15075100
Product Code LWQ
UDI-Device Identifier05414734006095
UDI-Public05414734006095
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number23AGFN-756
Device Catalogue Number23AGFN-756
Device Lot NumberC00003397
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/26/2023
Supplement Date Received by Manufacturer08/24/2023
Initial Report FDA Received Date07/18/2023
Supplement Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
-
-