MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT TIB IMP PS INSERT SZ 3.5 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT TIB IMP PS INSERT SZ 3.5 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED Back to Search Results
Model Number 02-022-35-3509
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Laxity (4526); Implant Pain (4561)
Date of Event 02/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.D10.Concomitants: 6550213- 02-020-11-0335 - truliant ps cem fem ps cem right sz 3.5, 6378843- 02-022-45-3525 - truliant tib fit tray cem sz 3.5f / 2.5t, ab6522- 1200-a - cemex rx 40g low viscosity, ab6715- 1200-i - cemex isoplastic 40g, ab2342- 1220-i - cemex isoplastic 20g, 4996798- 200-02-26 - three peg patella 26mm, 4000621092- a10012 - gps implant kit v2.
 
Event or Problem Description
As reported via legal documentation, a patient had right knee replacement surgery on (b)(6), 2021.They underwent right knee revision surgery on (b)(6), 2023, approximately 1 year 3 months after their initial procedure.(b)(6), 2023 op report states that there was obvious evidence of oxidation of the polyethylene with yellow discoloration.There was also pitting noted in both the medial and lateral mid portion of the polyethylene.There was no evidence of any loosening of the actual components.It was also noted that the polyethylene head of the patella had some lateral edge wear and upon further inspection, it appeared that the button was placed medial and it was a small button and there was still residual patella.The surgeon noted it looked like the patella during flexion and extension, the lateral femoral condyle was riding on the edge of the patella and was tracking abnormally resulting in abnormal patella wear on the lateral edge of the patella.The patient was brought to the recovery room in stable condition.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
Mfr #1038671-2024-05094.This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP PS INSERT SZ 3.5 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
3523782617
MDR Report Key17346970
Report Number1038671-2023-01703
Device Sequence Number6543851
Product Code JWH
UDI-Device Identifier10885862304346
UDI-Public10885862304346
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date10/12/2025
Device Model Number02-022-35-3509
Device Catalogue Number02-022-35-3509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2023
Supplement Date Received by Manufacturer12/26/2024
Initial Report FDA Received Date07/18/2023
Supplement Report FDA Received Date12/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0023-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-