• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC EVOLVE STEM 7.5MM +2; PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC EVOLVE STEM 7.5MM +2; PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER Back to Search Results
Model Number 496S275
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Component Missing (2306); Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is not available for evaluation as it remains implanted in the patient.If additional information becomes available, it will be provided on a supplemental report.Device remains implanted.
 
Event or Problem Description
According to the report, a packing error was identified whereby an incorrect implant size was erroneously packaged with labeling inconsistent with its actual size.As a consequence, a surgical delay of approximately 15-20 minutes occurred.
 
Additional Manufacturer Narrative
The complaint was confirmed, since the product was returned for evaluation and matches the alleged failure.Inspection of the received information/evidences are done or will be done, in the proceedings of the nc.If more information is provided, the case will be reassessed.
 
Event or Problem Description
According to the report, a packing error was identified whereby an incorrect implant size was erroneously packaged with labeling inconsistent with its actual size.As a consequence, a surgical delay of approximately 15-20 minutes occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLVE STEM 7.5MM +2
Common Device Name
PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17347305
Report Number3010667733-2023-00392
Device Sequence Number14962862
Product Code KWI
UDI-Device Identifier00840420185877
UDI-Public00840420185877
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K060731
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number496S275
Device Catalogue Number496S275
Device Lot Number1704691
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2023
Supplement Date Received by Manufacturer09/11/2023
Initial Report FDA Received Date07/18/2023
Supplement Report FDA Received Date09/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexFemale
-
-