• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) IDENTITY ADX XL DR; IMPLANTABLE PACEMAKER PULSE-GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) IDENTITY ADX XL DR; IMPLANTABLE PACEMAKER PULSE-GENERATOR Back to Search Results
Model Number 5386
Medical Device Problem Codes Failure to Disconnect (2541); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the patient presented for a routine generator replacement.During the procedure, the right ventricular set screw on the pacemaker could not be loosened due to a header anomaly.The right ventricular lead was capped and the pacemaker was explanted and replaced.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported event of could not untighten the ventricle setscrew to remove the lead was confirmed.The pacemaker was returned for analysis.Analysis revealed that calcified blood was filling the v-setscrew inset.The calcified blood was preventing the wrench from engaging the setscrew inset to untighten the setscrew.Wrenches cannot penetrate the calcified blood.The wrench will just strip the portions of the inset it comes in contact with.When the calcified blood was removed in the lab, a wrench could be fully inserted into the setscrew inset and the setscrew could be untightened normally, and the stuck lead removed.The blood came through holes punched through the septum by the torque wrench at time of implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDENTITY ADX XL DR
Common Device Name
IMPLANTABLE PACEMAKER PULSE-GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17347821
Report Number2017865-2023-36017
Device Sequence Number14963005
Product Code DXY
UDI-Device Identifier05414734006903
UDI-Public05414734006903
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2009
Device Model Number5386
Device Catalogue Number5386
Device Lot Number0002217103
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2023
Supplement Date Received by Manufacturer09/18/2023
Initial Report FDA Received Date07/18/2023
Supplement Report FDA Received Date09/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PASSIVE PLUS DX
Patient SexUnknown
-
-