ZIMMER SWITZERLAND MANUFACTURING GMBH AVENIR MÜLLER, STEM, STANDARD, UNCEMENTED, HA, 3, TAPER 12/14; COATED HIP FEMUR PROSTHESIS, MODULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870)
Date of Event 02/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: foreign - israel.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device is used for treatment.Medical records were not provided.With the available information, a definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that a patient underwent hip prostheses implantation approximately seven (7) years ago.About six and half (6,5) years later patient had a low energy fall and sustained a b2 periprosthetic fracture.The patient underwent revision of the femoral component.There was minimal bone ingrowth on the removed stem.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVENIR MÜLLER, STEM, STANDARD, UNCEMENTED, HA, 3, TAPER 12/14
Common Device Name
COATED HIP FEMUR PROSTHESIS, MODULAR
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17349072
Report Number0009613350-2023-00398
Device Sequence Number6079946
Product Code LZO
UDI-Device Identifier00889024590953
UDI-Public(01)00889024590953(17)200930(10)2821280
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K193030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/30/2020
Device Model NumberN/A
Device Catalogue Number01.06010.003
Device Lot Number2821280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2023
Initial Report FDA Received Date07/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age89 YR
Patient SexFemale
Patient Weight70 KG
-
-