• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE SHIM SZ5 7MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE SHIM SZ5 7MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Model Number 2545-00-743
Medical Device Problem Codes Break (1069); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4) depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device was found broken and with an overall worn appearance.No reported surgical delays or patient consequences.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary devices were reported for unknown reason.The device associated with this report was returned to depuy synthes for evaluation.Visual inspection of the returned item found it was broken in two pieces.Additionally, the device presents and overall worn appearance.Potential cause can be attributed to repetitive heavy usage over long years of service.The damage observed is consistent with overload of the material due to excessive force applied to the device repeated times.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune shim sz5 7mm would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SHIM SZ5 7MM
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND - 9616671
loughbeg, ringaskiddy co
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17349634
Report Number1818910-2023-14616
Device Sequence Number5065799
Product Code HWT
UDI-Device Identifier10603295134152
UDI-Public10603295134152
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2545-00-743
Device Catalogue Number254500743
Device Lot NumberMVMCBQ310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2023
Supplement Date Received by Manufacturer09/07/2023
Initial Report FDA Received Date07/19/2023
Supplement Report FDA Received Date09/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-