| Model Number |
5356 |
| Medical Device Problem Code |
Failure to Disconnect (2541)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a routine device exchange procedure, it was reported one of the ventricular lead was unable to be disconnected from the device header.The lead was cut, capped and replaced.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
D6a - date of implant: implant date is estimated to have occurred in 2008.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of could not untighten the ventricle setscrew to remove the lead was confirmed.Analysis revealed that calcified blood was filling the v-setscrew inset.The calcified blood was preventing the wrench from engaging the setscrew inset to untighten the setscrew.Wrenches cannot penetrate the calcified blood.The wrench will just strip the portions of the inset it comes in contact with.When the calcified blood was removed in the lab, the wrench could be fully inserted into the setscrew inset and the setscrew could be untightened normally, and the stuck lead removed.The blood came through holes punched through the septum by the torque wrench at time of implant.
|
| |
|