| Model Number |
AMSCO 400 STEAM STERILIZER |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the unit and found a large hole in the view port allowing water and steam to leak from the sterilizer.The technician replaced the viewport, tested the unit, confirmed it to be operating according to specifications, and returned it to service.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported a water leak from the generator of their amsco 400 steam sterilizer onto the floor.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The technician collected water samples to test the facility's incoming water and the results identified that three of the critical chemical attributes of water quality were above the maximum requirements for the carbon-steel steam generator as stated in the amsco 400 sterilizer operator manual.(table 5-1.Required feed water quality for carbon-steel steam generators).The user facility is responsible for ensuring that their incoming water supply remains within the carbon-steel steam generator's operating requirements.The amsco 400 operator manual states (3-1), "water quality - supplied must be within specifications.Improper water quality adversely affects equipment operation." the technician notified the customer of the water quality test results and that their water quality is not meeting the required operating conditions.The user facility has committed to making the necessary improvements to the facility's incoming water quality.No additional issues have been reported.
|
| |
|
Search Alerts/Recalls
|