| Model Number |
2.4MM S-TAK PERC INSERT DISP KIT |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event has not yet been determined as the device has not been returned for evaluation at this time.A follow-up report will be submitted, including a most likely cause if a root cause can not be determined.
|
| |
|
Event or Problem Description
|
|
On 06/20/2023, it was reported by a arthrex employee via sems that an ar-1934pi perc insert kit drill bit bent and became stuck inside the guide.This was discovered during a case with no patient effect.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to misuse due to misaligned insertion; prying/leveraging the device during insertion.
|
| |
|