• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB Back to Search Results
Model Number TV-IL1422120-J
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 06/26/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.H3 other text : devices remain implanted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of the alto abdominal aortic aneurysm stent system.During the initial implant procedure, angiography identified bilateral type ib endoleaks.Additional balloon touch-up was performed, however the endoleak remained unresolved.In response, the physician occluded the right internal iliac artery and an additional limb was added on the right-side.Balloon touch-up was performed again, and angiography showed a type iiia endoleak between the two right limbs and the persistent type ib endoleak from the left limb.The physician elected to conclude the procedure and will continue to monitor the patient.Additional treatment is being discussed, but an additional surgery has not been scheduled.The patient is reportedly in stable condition.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported ovation ix, type iiia endoleak (between the right common iliac limbs) is unconfirmed.The reported bilateral type ib endoleaks (right-side resolved) are confirmed.This is moderately consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: h6 investigation finding codes; remove 3233.H6 investigation finding codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
ILIAC LIMB
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer (Section G)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
,
santa rosa CA 95403
Manufacturer Contact
gary kirchgater
3910 brickway blvd
,
santa rosa, CA 95403
8009832284
MDR Report Key17352785
Report Number3008011247-2023-00144
Device Sequence Number12068804
Product Code MIH
UDI-Device Identifier00850007370336
UDI-Public(01)00850007370336(17)250609(10)FS060822-56
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/09/2025
Device Model NumberTV-IL1422120-J
Device Catalogue NumberTV-IL1422120-J
Device Lot NumberFS060822-56
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/27/2023
Supplement Date Received by Manufacturer08/08/2023
Initial Report FDA Received Date07/19/2023
Supplement Report FDA Received Date09/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTO MAIN BODY, LN FS092222-28; OVATION IX ILIAC LIMB, LN FS071422-26; OVATION IX ILIAC LIMB, LN FS090922-52; OVATION PRIME FILL POLYMER, LN FF081822-01
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-