• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INTRAVIA CONTAINER EMPTY; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INTRAVIA CONTAINER EMPTY; CONTAINER, I.V. Back to Search Results
Catalog Number 2J8002
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stopper of a 250ml non-dehp fluid path intravia container completely came out from the injection port with the needle.The issue was identified during compounding, when after the first needle stick.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which observed roughness and separation of the injection site; however, the other components are place according to the specification.Due to the nature of the returned sample no additional testing could be performed.The reported condition was verified, however, the cause could not be determined from the photograph.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA CONTAINER EMPTY
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17353132
Report Number1416980-2023-03662
Device Sequence Number17509585
Product Code KPE
UDI-Device Identifier00085412029573
UDI-Public(01)00085412029573
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K964853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2J8002
Device Lot NumberDR22C07048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date07/19/2023
Supplement Report FDA Received Date08/23/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-