| Model Number |
CI-1600-05 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
During implantation surgery, the electrode was partially inserted.Intra-operative imaging confirmed partial insertion.The recipient reportedly experienced intermittent connection during initial activation of the device.X-ray imagining showed the electrode array outside of the cochlea.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics device.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|