• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; IMPLANTABLE DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; IMPLANTABLE DEFIBRILLATOR Back to Search Results
Model Number N160
Medical Device Problem Codes High impedance (1291); Pacing Problem (1439)
Health Effect - Clinical Codes Arrhythmia (1721); Electric Shock (2554)
Date of Event 06/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range shock impedance measurements.It was decided to address the issue at the next device replacement.Additionally, this patient experienced episodes of pacemaker mediated tachycardia (pmt).The device's algorithm successfully terminated the episode of pmt.During the replacement procedure a defibrillation threshold (dft) test was performed in order to evaluate the measurements, although they were trending down, they were still out of range.This device was explanted and replaced.The patient will continue to be monitored.No further adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a malfunction or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a malfunction or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range shock impedance measurements.It was decided to address the issue at the next device replacement.Additionally, this patient experienced episodes of pacemaker mediated tachycardia (pmt).The device's algorithm successfully terminated the episode of pmt.During the replacement procedure a defibrillation threshold (dft) test was performed in order to evaluate the measurements, although they were trending down, they were still out of range.This device was explanted and replaced.The patient will continue to be monitored.No further adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
IMPLANTABLE DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17356576
Report Number2124215-2023-38241
Device Sequence Number20337541
Product Code LWP
UDI-Device Identifier00802526480997
UDI-Public00802526480997
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/27/2014
Device Model NumberN160
Device Catalogue NumberN160
Device Lot Number881820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/29/2023
Supplement Date Received by Manufacturer07/28/2023
Initial Report FDA Received Date07/19/2023
Supplement Report FDA Received Date08/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/29/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-