| Catalog Number |
09180443001 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of a software issue for the cobas infinity core license.The reporter stated that they have a rule to delete a specific test (hiv2) when a specific result (non-reactive) was received.This rule is triggered when a result is received from the instrument.The reporter stated that for one case, this rule was not triggered and the test was not deleted.The reporter detected the issue when the result was incorrectly exported to the laboratory information system (lis).
|
| |
|
Additional Manufacturer Narrative
|
|
The software version is 3.03.08.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|