Section e3: occupation is patient/consumer (patient's wife) on a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." the investigation did not identify a product problem.The cause of the event could not be determined.H3 other text : na.
|
|
There was an allegation of questionable results from coaguchek xs meter, serial number (b)(6), compared to a laboratory result using an unknown method.At 7:00 a.M., the laboratory result was 2.5 inr.At 9:00 a.M., the meter result was 1.5 inr.No medication changes were made based on the meter result.The patient¿s therapeutic range is 2.5 - 3.0 inr.The patient's testing frequency is weekly.
|