| Model Number |
D400 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range pacing impedance measurements.It was decided to turn off the therapy and provide an external defibrillator device.This device remains in service.No further adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range pacing impedance measurements.It was decided to turn off the therapy and provide an external defibrillator device.This device remains in service.No further adverse patient effects were reported.Additional information was received detailing that this device was explanted and replaced.This device was returned to boston scientific for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to the following fields:
b2: outcome attributed to adverse event,
b5: describe event or problem field,
d6b: explant date,
h6: impact codes.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range pacing impedance measurements.It was decided to turn off the therapy and provide an external defibrillator device.This device remains in service.No further adverse patient effects were reported.Additional information was received detailing that this device was explanted and replaced.This device was returned to boston scientific for analysis.
|
| |
|