EXACTECH, INC. LOGIC CR TIB INSERT SLOPE++, SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
LOGIC CR TIB INSERT SLOPE++, SZ 3, 9MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
06/29/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products.Logic cr femoral por, left, sz 3 (cat# 02-010-04-0230 / serial#(b)(6)).Three peg patella 35mm (cat# 200-02-35 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately four years post initial left side tka, the female patient had a revision due to poly wear.Patient was revised into a logic ps tkr.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Images received.Sales rep was unable to obtain x-rays.The devices are not available for evaluation as they were disposed at the hospital, no other patient information/medical history reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the revision reported was likely the result of improper implant position, the overall posterior slope of the insert and/or potential ligamentous/soft tissue imbalances which allowed for increased posterior contact between the femoral component and the tibial insert and led to the observed damage of the polyethylene.Additionally, a contributing factor to the observed damage may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device was not returned for evaluation and radiographs were not provided.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported was likely the result of improper implant position, the overall posterior slope of the insert and/or potential ligamentous/soft tissue imbalances which allowed for increased posterior contact between the femoral component and the tibial insert and led to the observed damage of the polyethylene.Additionally, a contributing factor to the observed damage may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device was not returned for evaluation and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|