| Model Number |
VAL KREULOCK SCREW, TI, 3.0X18 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event has not yet been determined as the device has not been returned for evaluation at this time.A follow-up report will be submitted, including a most likely cause if a root cause can not be determined.
|
| |
|
Event or Problem Description
|
|
On 06/14/2023, it was reported by a sales representative via phone that an ar-8933vcl-18 kreulock screw would not thread in the plate, it would just spin.This occurred during the first mtp fusion on (b)(6) 2023, a hole was drilled and when they tried to insert the screw it would not thread it was just spinning.The screw was removed, and the case was completed using another screw.There was no effect on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misaligned insertion and/or prying/leveraging the screw during insertion.
|
| |
|