• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER INSERTION TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER INSERTION TRAY Back to Search Results
Model Number 802010
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that patient stated that one of the items was punctured.The head had a puncture through the tap.Per follow up via email received on (b)(6)2023, customer stated that no issue with a patent.The product arrived with a puncture in the package.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample were confirmed to exhibit the reported failure.Visual evaluation of the returned photo sample noted one unopened (with original packaging), bardia insertion tray.Visual inspection of the photo sample noted a hole puncture in the package.This does not meet specification which states "no damage or holes on trays or tyvek are allowed".A potential root cause for this failure mode could be ¿material handling." the lot number is unknown; therefore, the device history record could not be reviewed.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that patient stated that one of the items was punctured.The head had a puncture through the tap.Per follow up via email received on 11jul2023, customer stated that no issue with a patent.The product arrived with a puncture in the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER INSERTION TRAY
Common Device Name
FOLEY CATHETER INSERTION TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17364153
Report Number1018233-2023-05348
Device Sequence Number17449635
Product Code OHR
UDI-Device Identifier00801741018282
UDI-Public(01)00801741018282
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number802010
Device Catalogue Number802010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2023
Supplement Date Received by Manufacturer10/16/2023
Initial Report FDA Received Date07/20/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-