| Model Number |
119108 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that no outflow of urine after catheter insertion.When the catheter was removed and water was injected, it could not be drained.Possible blockage at the tip of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed because the reported failure could not be reproduced.The reported failure is within specification as the reported failure could not be reproduced.It is unknown if the product was used for patient treatment or diagnosis.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.Visual evaluation of the returned sample noted one opened (with original packaging), all-silicone temp sensing foley catheter.Visual inspection of the sample noted no obvious visible defects.The catheter balloon was inflated with 3.5ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated in under 2 minutes with no cuffing or leaks noted, returning 3.5ml of solution.This is within specification per inspection procedure, which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The labeling review was not required as the reported event was unconfirmed.Correction: d h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that no outflow of urine after catheter insertion.When the catheter was removed and water was injected, it could not be drained.Possible blockage at the tip of the catheter.
|
| |
|
Search Alerts/Recalls
|